XU Bo,ZHANG Zhongmin,ZHAO Weidong.Pressure variation of facet joint following cervical total disc replacement anterior cervical discectomy and fixation[J].Chinese Journal of Spine and Spinal Cord,2010,20(5):406-410.
Pressure variation of facet joint following cervical total disc replacement anterior cervical discectomy and fixation
Received:February 11, 2010  Revised:March 24, 2010
English Keywords:Cervical spine  Cervical total disc replacement  Fixation  Pressure sensitive film  Biomechanics
Fund:
Author NameAffiliation
XU Bo Department of OrthopedicsThe Third Affiliated Hospital of Southern Medical UniversityGuangzhou510630China 
ZHANG Zhongmin 南方医科大学第三附属医院骨科 510630 广州市 
ZHAO Weidong 南方医科大学医用生物力学实验室 510515 广州市 
朱青安 南方医科大学南方医院骨科 510515 广州市 
周荣平 南方医科大学第三附属医院骨科 510630 广州市 
金大地 南方医科大学第三附属医院骨科 510630 广州市 
Hits: 4421
Download times: 3554
English Abstract:
  【Abstract】 Objective:To measure the pressure variation of facet joint after cervical total disc replacement(TDR) anterior cervical discectomy and fixation(ACDF),and explore mechanical effect for cervical facet joint following two operations.Method:A total of 6 adult human fresh cervical spinal specimens(C2-C7) were divided into intact cervical specimens group,TDR group and ACDF group.Following intact specimens analysis,all specimens were sequentially reconstructed at C4-C5 segment with TDR and ACDF.Pressure sensitive film was put into left-side facet joints of C4-C5 segment.Firstly a 75N follower load was applied,then a three-dimensional movement was performed with the continous load.Pressures of facet joint were measured in 0N·m and 2.0N·m(flexion/extension,left/right lateral bending),4.0N·m moment(left/right axial rotation).Result:To compare with the pressure of left-side facet joint in neutral position,pressure of facet joint of intact specimens and TDR increased significantly in extension,left lateral bending and right axial rotation (P<0.05),but not obviously changed in flexion,right lateral bending and left axial rotation.In contrast with intact specimens and TDR,the pressure of facet joint of ACDF were not significant between in neutral position and any motions(P>0.05).In the comparison among three tested groups,the pressure variation of facet joint was similar between intact specimens and TDR.The pressures of facet joint of ACDF were obviously decreased in extension,left lateral bending and right axial rotation,but there were not significance among three groups(P>0.05).Conclusion:Pressure variation of facet joint is not significant between TDR and intact specimens.It indicates that cervical total disc replacement can effectively restore normal physiological pressure of cervical facet joint.