王 迪,卫 磊,雷 伟,吴子祥.新型翻修延长棒的设计及其在翻修模型中的生物力学研究[J].中国脊柱脊髓杂志,2017,(11):1019-1023.
新型翻修延长棒的设计及其在翻修模型中的生物力学研究
Design and biomechanical evaluation of a novel connective rod for revision surgery
投稿时间:2017-07-20  修订日期:2017-10-22
DOI:
中文关键词:  翻修手术  模型验证  脊柱融合术  生物力学
英文关键词:Reoperation  Model validation  Spinal fusion  Biomechanics
基金项目:国家自然科学基金资助项目(81301535),陕西省重点研发计划(2017SF-024)
作者单位
王 迪 第四军医大学附属西京医院骨科 710032 西安市 
卫 磊 第四军医大学附属西京医院骨科 710032 西安市 
雷 伟 第四军医大学附属西京医院骨科 710032 西安市 
吴子祥  
摘要点击次数: 2698
全文下载次数: 1238
中文摘要:
  【摘要】 目的:采用对端结合的方式,设计适用于脊柱翻修手术的延长棒,研究其生物力学特性。方法:依据美国材料与试验协会(American Society for Testing and Materials,ASTM)的标准测试方法,制备椎体切除术后腰椎模型。采用静态压缩弯曲试验、静态拉伸弯曲试验和疲劳试验,分别测量“传统连接棒”(A组,n=6)和“翻修延长棒”(B组,n=6)的屈服载荷、弯曲刚度和疲劳参考值,初步评价“翻修延长棒”的生物力学稳定型。结果:在静态压缩弯曲试验中,A组试验模型屈服载荷为865.60±25.71N,弯曲刚度为52.35±0.38N/mm。B组试验模型屈服载荷为863.20±11.01N,弯曲刚度为52.92±2.43N/mm,两组之间无统计学差异(P>0.05)。在静态拉伸弯曲试验中,A组试验模型屈服载荷为1184.20±30.14N,拉伸弯曲刚度为58.89±3.01N/mm;B组试验模型屈服载荷为1169.20±14.89N,拉伸弯曲刚度为57.48±1.47N/mm,与A组比较均无统计学差异(P>0.05)。在疲劳试验中,A组的疲劳载荷参考值为303N,B组的疲劳载荷参考值为302N,与A组比较无统计学差异(P>0.05)。结论:新型翻修延长棒具有与传统连接棒相似的屈服载荷、弯曲刚度及耐疲劳性。
英文摘要:
  【Abstract】 Objectives: To design a novel connective rod for revision surgery by using an end to end connection joint, and to evaluate their biomechanical characters. Methods: Following the guidelines of American society for testing and materials(ASTM), static compression bending test, static tension bending test, fatigue test was employed on both conventional rod(group A, n=6) and connective rod for revision surgery(group B, n=6) to measure the bending yield load, bending stiffness and maximum run-out load, and to determine the biomechanical stability of novel connective rod. Results: In static compression bending test, compressive bending yield load and compressive bending stiffness of group A were 865.60±25.71N and 52.35±0.38N/mm respectively; while they were 863.20±11.01N and 52.92±2.43N/mm in group B, the differences between groups had no statistical significance. In static tension bending test, tensile bending yield load and tensile bending stiffness of group A were 1184.20±30.14N and 58.89±3.01N/mm respectively; and for group B, they were 1169.20±14.89N and 57.48±1.47N/mm respectively, the differences between groups had no statistical significance. In fatigue test, the maximum run-out load of group A and group B was 303N and 302N respectively, the difference between groups had no statistical significance. Conclusions: Compared with the conventional connective rod, the novel connective rod for revision surgery has a similar bending yield load, bending stiffness and maximum run-out load.
查看全文  查看/发表评论  下载PDF阅读器
关闭
function PdfOpen(url){ var win="toolbar=no,location=no,directories=no,status=yes,menubar=yes,scrollbars=yes,resizable=yes"; window.open(url,"",win); } function openWin(url,w,h){ var win="toolbar=no,location=no,directories=no,status=no,menubar=no,scrollbars=yes,resizable=no,width=" + w + ",height=" + h; controlWindow=window.open(url,"",win); } &et=52A59D2F0854D9F4B80FBA59CED70C02355861896E9164983982FCDAF7663EA893E76A98B6EE5F6CF3CABA604DC0C7725BDA4BBF3F8E8D4E19FA92D9A0393679277D6863DFE1B942ADFDA72ED900A3B5956EB11302A18BAD&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=527A01A248DACB72&jid=9262A66F274A6CFEEBE23EC40CDE54FC&yid=FA004A8A4ED1540B&aid=E922C88C1973C4B3F0601CEA9CD2490B&vid=&iid=708DD6B15D2464E8&sid=CBC69BEA05C12902&eid=A40901B6135B333E&fileno=20171110&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="527A01A248DACB72"; var my_jid="9262A66F274A6CFEEBE23EC40CDE54FC"; var my_yid="FA004A8A4ED1540B"; var my_aid="E922C88C1973C4B3F0601CEA9CD2490B";