蒋伟宇,于 亮,马维虎,赵刘军,徐荣明.寰枢椎后路动态固定系统有效性的实验研究[J].中国脊柱脊髓杂志,2013,(12):1097-1100.
寰枢椎后路动态固定系统有效性的实验研究
The efficacy of posterior atlanto-axial dynamic fixation system, an experimental study
投稿时间:2013-04-18  修订日期:2013-07-30
DOI:
中文关键词:  寰枢椎  动态固定系统  有效性  生物力学
英文关键词:Atlanto-axial  Dynamic fixation system  Effectiveness  Biomechanics
基金项目:浙江省医药卫生科技计划项目(编号:2011RCB034);浙江省自然科学基金项目(编号:LY02H06001)
作者单位
蒋伟宇 浙江省宁波市第六医院脊柱外科 315040 
于 亮 浙江省宁波市第六医院脊柱外科 315040 
马维虎 浙江省宁波市第六医院脊柱外科 315040 
赵刘军  
徐荣明  
摘要点击次数: 3294
全文下载次数: 2106
中文摘要:
  【摘要】 目的:测试自主设计的寰枢椎后路动态固定系统的有效性。方法:根据寰枢椎的解剖特征及常规寰枢椎后路椎弓根螺钉置钉技术,设计符合寰枢椎运动方式的新型寰枢椎后路动态固定系统,并测试其有效性。将8具成人枕颈标本(C0~C4)依次按完整状态组(A组)、寰枢椎不稳状态组(B组)、寰枢椎后路动态固定状态组(C组)及寰枢椎后路椎弓根螺钉固定状态组(D组)的顺序进行测试。在三维运动试验机(KD-101)上测试4组模型的颈椎运动范围(range of motion,ROM),所有标本测试时施加1.50N·M力偶矩按前屈/后伸、左/右侧屈、左/右轴向旋转的顺序进行,取其均值,所得结果进行比较。结果:B组与A组在屈伸状态、侧屈状态和轴向旋转状态的ROM分别相差41%、44%、9%,差异有统计学意义(P<0.05)。C组与A组在屈伸、侧屈和轴向旋转状态的ROM分别相差8%、9%、9%,差异无统计学意义(P>0.05);C组ROM与B组比较分别相差46%、50%、17%,差异有统计学意义(P<0.001)。D组固定后在屈伸、侧屈、轴向旋转状态的ROM与A组分别相差35%、46%、58%,差异有统计学意义(P<0.001);与B组分别相差62%、70%、61%,差异有统计学意义(P<0.001);与C组分别相差30%、41%、53%,差异有统计学意义(P<0.001)。结论:寰枢椎后路动态固定系统能够有效地限制寰枢椎的不稳,同时又能保留寰枢椎一定的屈伸、侧屈和旋转活动功能。
英文摘要:
  【Abstract】 Objectives: To test the efficacy of self-designed posterior dynamic fixation system. Methods: According to the anatomic features and routine posterior pedicle screw technique of atlanto-axial vertebrae, a posterior atlanto-axial dynamic fixation system was designed, and its effectiveness was tested. Eight human cadaveric specimens(C0-C4) under test were subgrouped as: the intact group(group A), instable group(group B), posterior atlanto-axial dynamic fixation group(group C) and posterior atlanto-axial pedicle screw fixation group(group D). The three dimensional ranges of motions of C1 relative to C2 under loading of ±1.5N·M were measured under flexion/extension, left and right lateral bending, and left and right axial rotation. The obtained data were statistically analyzed. Results: The results showed that: (1)the difference between group B and A in ROM of flexion/extension, left/right lateral bending and rotation was 41%, 44% and 9%, respectively, with significant differences(P<0.05); the difference between group C and B in ROM was 46%, 50% and 17%, respectively, with significant differences(P<0.001); the difference between group D and B in ROM was 62%, 70% and 61%, respectively, with significant differences(P<0.001). (2)the difference between group C and A in ROM of flexion/extension, left/right lateral bending and rotation was 8%, 9% and 9%, respectively, with no significant differences(P>0.05). (3)the difference between group D and A in ROM of flexion/extension, left/right lateral bending and rotation was 35%, 46% and 58%, respectively, with significant differences(P<0.001); and compared with group C, the difference was 30%, 41% and 53% respectively, which showed significant differences(P<0.001). Conclusions: The posterior atlanto-axial dynamic fixation is effective for atlanto-axial dislocation and instability, as well as maintaining the ROM of C1-C2.
查看全文  查看/发表评论  下载PDF阅读器
关闭
function PdfOpen(url){ var win="toolbar=no,location=no,directories=no,status=yes,menubar=yes,scrollbars=yes,resizable=yes"; window.open(url,"",win); } function openWin(url,w,h){ var win="toolbar=no,location=no,directories=no,status=no,menubar=no,scrollbars=yes,resizable=no,width=" + w + ",height=" + h; controlWindow=window.open(url,"",win); } &et=05B7EC569A6F2359E8507AD8D4B0D73BE5BCE1471610497588DFD27CDE9FFB3B61486B11711FD9FB70E151EAAE58B18477F9A4B0409286A017FE5A0B2F30D44293A1252FBCB4D28E42BD74E66AB25A549D4B056DD8BDBC30&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=527A01A248DACB72&jid=9262A66F274A6CFEEBE23EC40CDE54FC&yid=FF7AA908D58E97FA&aid=524FCDFE05EF299076E3A55165B7DAC3&vid=&iid=59906B3B2830C2C5&sid=FB3C6F66BC48DD45&eid=D7B9DD9E919B6180&fileno=20131208&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="527A01A248DACB72"; var my_jid="9262A66F274A6CFEEBE23EC40CDE54FC"; var my_yid="FF7AA908D58E97FA"; var my_aid="524FCDFE05EF299076E3A55165B7DAC3";